Automation

Differential pressure accuracy

The error of a differential pressure measurement is fixed in pressure units by the sensor's upper range limit, not by the span you calibrate. This page shows how the tool combines reference accuracy and thermal effect for an electronic dP (two transmitters) and for a conventional single-cell dP, and how that error propagates to level, density and flow — where the square root and the 1/Q² growth change the answer completely.

Calculate now

Free, no sign-up. Runs in your browser — nothing you type leaves this page.

When to use

When choosing between an electronic dP and a conventional transmitter with capillaries; when checking whether a level, interface, density or flow loop meets the accuracy in the project specification; when deciding what actually pays — a better accuracy class, a lower URL or a narrower temperature range; and when building the uncertainty budget of an orifice plate per ISO 5167-2.

The error lives in pressure units, set by the URL

A pressure transmitter’s datasheet states its accuracy as a percentage, but the physics behind it is an absolute error in pressure, fixed by the sensor’s upper range limit. A 2000 mbar sensor declared at 0.1 % of URL errs by 2 mbar whether it is calibrated for 2000 mbar or for 200 mbar. Calibrating a narrower span does not shrink the error; it only makes the same error a larger percentage.

The tool therefore works internally in pascals, converts every input from the one global pressure unit chosen on screen, and only converts back at the end. A single unit for all fields is deliberate: mixing bar and mbar between URL and span is the classic thousand-fold mistake.

Electronic dP against the conventional cell

In an electronic dP two independent transmitters sit on the high- and low-pressure taps and the electronics subtract them. Each one errs on its own URL, and because both must be ranged for the full static pressure, that URL is typically much larger than the differential span. The two errors combine — by RSS or linear sum — and the tool reports which sensor dominates. When the contributions are balanced, improving only one sensor buys little.

In a conventional dP a single cell sees only the differential. The base accuracy is declared on the calibrated span, but most datasheets add a turndown rule of the form A[%span] = a + k·TD above a stated limit. Rewritten in pressure units that is a·span + k·URL, and the tool takes the larger of the base accuracy and the rule. Its thermal effect uses the two-term form, one part on URL and one on span. Either way the thermal term is scaled by m = 1 in band mode, or by |T_process − T_ref|/ΔT_ref when the coefficient is declared per degree interval.

From pressure error to level, density and flow

The pressure error only becomes a decision once it is expressed in the measured quantity. For level, h = ΔP/(ρ·g): the error in millimetres is the same at the bottom and at the top of the range, and the liquid density uncertainty enters one for one on top of it. (When a non-contact sensor is chosen instead, the question becomes geometric — see radar level transmitter mounting position.) For density, ρ = ΔP/(g·H): the divisor g·H is small, because H is typically 1–3 m, which makes this the most demanding application — and the static head of the fluid often suppresses a zero far larger than the useful range.

For flow, Q ∝ √ΔP halves the relative error, but ΔP falls with the square of flow. With a conventional cell of 1000 mbar URL calibrated to 250 mbar (0.1 % of span base accuracy, thermal 0.05 % URL + 0.1 % span), the total is 0.79 mbar. Across the range, on an orifice plate with β = 0.65 under ISO 5167-2, the combined flow uncertainty goes 0.69 % at 100 % of Qmax, 0.74 % at 70 %, 0.92 % at 50 %, 1.60 % at 33 % and 2.62 % at 25 %. The transmitter part alone grows from 0.16 % to 2.53 % — the 1/Q² law in numbers. A magnetic flow meter has no square root in its signal, which is why it is the usual alternative when the turndown is wide.

The ISO 5167-2 mode

For an orifice plate (the orifice plate lesson covers the device itself) the tool can break the flow uncertainty into the terms of ABNT NBR ISO 5167-2:2011. u(C) comes from clause 5.3.3.1 as a function of β, with the two arithmetic additions the standard specifies for small pipes and low Reynolds; u(ε) is 3.5·ΔP/(κ·p₁) for gas and exactly zero for liquid, evaluated point by point because it also varies with ΔP. The bore and pipe diameter uncertainties are weighted by sensitivity coefficients derived from the flow equation, 2/(1−β⁴) and 2β⁴/(1−β⁴).

The limits of clause 5.3.1 — d ≥ 12.5 mm, 50 ≤ D ≤ 1000 mm, 0.10 ≤ β ≤ 0.75 and the Reynolds minimum that depends on the tap type — are checked explicitly. Outside them the standard’s uncertainty does not apply and the tool raises an error; with Re_D left blank it reports applicability as not assessed rather than as passed. If u(C) is missing, the budget is incomplete and the verdict is undefined, following the GUM: an unknown term is not a zero term.

What this tool does not cover

The result is reference accuracy plus thermal effect. Static pressure effect, long-term stability, mounting position, power supply effect and, on an electronic dP, the static head error between taps at different elevations are outside it, as are wet-leg and seal effects and the uncertainty of the tap separation H in density service. The tool embeds no manufacturer data: every coefficient is read by the engineer from the instrument datasheet. In flow it does not size the primary element or compute the discharge coefficient itself — only its uncertainty; for that, follow how to size an orifice plate step by step.

Formulas and fundamentals

Electronic dP (two sensors) A_Diff = comp(A_HP·URL_HP/100, A_LP·URL_LP/100) | T_Diff = comp(T_HP·URL_HP/100, T_LP·URL_LP/100) · m

comp is RSS, √(Σx²), or the linear sum Σx for a worst case. The difference is made by subtraction in the electronics, so the error of each sensor scales with its own URL — never with the differential span.

Conventional dP (single cell) with turndown rule TD = URL/span | A_base = a_base·span | A_td = a·span + k·URL (if TD > limit) | A_Diff = max(A_base, A_td)

The datasheet form A[%span] = a + k·TD rewritten in pressure units. Above the stated turndown the rule governs and the error becomes proportional to the cell URL. In "direct" mode the accuracy at the calibrated span is entered ready and no rule is applied.

Single-cell thermal effect T_Diff = (t_URL·URL + t_span·span) · m

The universal two-term form for a single cell: a part proportional to the URL (zero shift) and a part proportional to the span (span shift).

Thermal multiplier band: m = 1 | ΔT mode: m = |T_process − T_ref| / ΔT_ref

In band mode the coefficient is already closed for the temperature range on the datasheet; in ΔT mode a coefficient declared "per 28 °C" or "per 10 K" is scaled by the actual departure from the reference temperature.

Total and share of each term E_total = comp(A_Diff, T_Diff) | share_T = T_Diff² / (A_Diff² + T_Diff²)

The share (quadratic in RSS, linear in the sum) tells where the money goes: when the thermal term dominates, a better reference class buys almost nothing.

Propagation to the measured quantity level: δh = E/(ρ·g) | density: δρ = E/(g·H) | flow: δQ/Q = ½·E/ΔP, ΔP = span·(Q/Qmax)²

Level and density have sensitivity 1 and a constant absolute error across the range. Flow has sensitivity ½, but ΔP falls with the square of flow, so the relative flow error grows as 1/Q². g = 9.80665 m/s².

ISO 5167-2 orifice budget u(Q) = comp(u(C), u(ε), 2/(1−β⁴)·u(d), 2β⁴/(1−β⁴)·u(D), ½·E/ΔP, ½·u(ρ))

u(C) per clause 5.3.3.1: (0.7 − β) % for β < 0.20, 0.5 % for 0.20–0.60, (1.667β − 0.5) % above 0.60, plus arithmetic additions for D < 71.12 mm and for β > 0.5 with Re_D < 10 000. u(ε) = 3.5·ΔP/(κ·p₁) % for gas, zero for liquid.

Standards & methods

  • ABNT NBR ISO 5167-2:2011 — orifice plates; clause 5.3.1 (limits of use) and 5.3.3 (uncertainty of C and ε)
  • ISO 5167-1 — general principles of flow measurement by differential pressure devices
  • JCGM 100:2008 (GUM) — combination of uncertainties; an unknown term is not a zero term
  • Instrument datasheets — every accuracy, thermal and turndown coefficient is an input read from the manufacturer's data, none is built in

Typical reference values

Quantity Typical range Note
Reference accuracy of a modern dP transmitter 0.04 – 0.1 % of span usually degrades above a stated turndown
Turndown where the accuracy rule usually kicks in 5:1 – 10:1 read it from the datasheet; the tool applies only what is entered
Error growth in flow vs. full scale 4× at 50 % · 16× at 25 % of Qmax transmitter term only, from ΔP ∝ Q²
u(C) of an orifice plate, 0.20 ≤ β ≤ 0.60 0.5 % ABNT NBR ISO 5167-2:2011, 5.3.3.1
Applicability of ISO 5167-2 orifice plates d ≥ 12.5 mm · 50 ≤ D ≤ 1000 mm · 0.10 ≤ β ≤ 0.75 Re_D ≥ 5000, and ≥ 16 000·β² for β > 0.56 (corner and D-D/2 taps)
Level error per 1 mbar at ρ = 1000 kg/m³ ≈ 10.2 mm δh = 100 Pa / (1000 · 9.80665)

Worked example

Level in a pressurized vessel with an electronic dP (LT-01)

Inputs

Arrangement
Electronic dP (2 sensors) —
URL of each sensor
2000 mbar
Reference accuracy, HP and LP
0.1 % URL
Thermal effect, HP and LP (band)
0.2 % URL
Calibrated span
0 – 500 mbar
Liquid density
900 kg/m³
Requirement
2 % of span

Results

Reference accuracy term A_Diff
2.83 mbar
Thermal term T_Diff
5.66 mbar
Total error (RSS)
6.32 mbar
Total in % of span
1.26 %
Level error / level span
± 71.7 / 5665 mm
Turndown URL/span
4.0 :1
Thermal share of the error
80 %
Verdict
Meets 1.26 % ≤ 2 %

The loop meets the 2 % requirement with 1.26 % of span, which is ±72 mm of level — constant along the whole 5.67 m range, so near the bottom it is a much larger fraction of what is in the vessel. The decisive number is the 80 % thermal share: the reference accuracy contributes only 2.83 mbar. To improve this loop, narrow the process temperature range or pick lower-URL sensors; buying a better accuracy class is close to useless. Static pressure, long-term stability, mounting position and the density shift of the liquid are outside this figure and must fit in the remaining margin.

Run your own calculation Create a free account and get this for your own inputs in seconds.

Common mistakes

  • Applying the % of an electronic dP to the differential span. Each sensor errs on its own URL: two 2 bar sensors measuring 500 mbar carry four times the error a reading of "% of span" suggests.
  • Buying a better accuracy class when the thermal term dominates. In the worked example 80 % of the error is thermal — doubling the reference accuracy moves the total from 1.26 % to about 1.17 % of span.
  • Calibrating a wide-range cell to a narrow span and keeping the base accuracy. Above the stated turndown the datasheet rule replaces it, and the error becomes proportional to the URL.
  • Quoting a flow accuracy at full scale only. With ΔP ∝ Q², the transmitter contribution at 25 % of Qmax is 16 times larger than at 100 %.
  • Treating an unknown uncertainty as zero. Without u(C) the RSS loses its dominant term and the number improves — the tool reports the verdict as undefined instead.
  • Leaving Re_D blank and reading "within ISO 5167-2". Without Reynolds the standard's applicability is not assessed at all, and the tool says so.

Frequently asked questions

Why is an electronic dP less accurate than a conventional transmitter on small spans?

Because each sensor errs on its own URL and both errors add. The two sensors must be ranged for the full static pressure plus the differential, so their URL is far larger than the span. A conventional cell sees only the differential and errs, at most, on its own URL. The electronic dP wins on response time and on eliminating capillaries, not on accuracy at small spans.

Should I combine the terms by RSS or by linear sum?

RSS assumes independent random terms and is the usual "total probable error" of datasheets. The linear sum is a worst case — every term at its limit with the same sign. The tool offers both; RSS is the default.

How much does turndown hurt?

In pressure units the error is fixed by the URL. Narrowing the span without lowering the URL raises the error in % of span in the same proportion. In a single cell, beyond the stated turndown the manufacturer rule A = a + k·TD replaces the base accuracy.

Why does flow accuracy get so much worse at low flow?

Q is proportional to √ΔP, which halves the relative error (sensitivity ½), but ΔP falls with the square of flow. At 50 % of Qmax the transmitter term is 4 times its full-scale value; at 25 %, 16 times. This is the physical reason for the practical rangeability limit of differential producers.

What does β change in the orifice uncertainty?

The bore uncertainty enters with weight 2/(1−β⁴) and the pipe uncertainty with 2β⁴/(1−β⁴). At β = 0.65 those weights are 2.43 and 0.43: gauging the bore precisely pays far more than gauging the pipe. β also sets u(C) itself, which rises from 0.5 % to (1.667β − 0.5) % above β = 0.60.

Does the result include static pressure effect and stability?

No. The tool computes reference accuracy plus thermal effect only. Static pressure effect, long-term stability, mounting position, supply effect and, on an electronic dP, the static head between taps at different elevations are outside, and the tool states this scope in every run.

Glossary

URL (upper range limit)
The largest pressure the sensor can be calibrated to. Errors declared in % of URL scale with it, not with the calibrated span.
Span
Difference between the upper and lower calibrated values (URV − LRV).
Turndown (TD)
Ratio URL/span — how far the sensor has been ranged down.
Electronic dP (EdP)
Two pressure sensors, one on each tap, with the differential computed by subtraction in the electronics instead of in a single cell.
Reference accuracy
Error of the transmitter at reference conditions — linearity, hysteresis and repeatability — as declared in the datasheet.
Thermal effect
Shift of zero and span caused by ambient or process temperature departing from the calibration reference.
RSS
Root sum of squares, the combination of independent uncertainties used for the "total probable error".
Sensitivity coefficient
Partial derivative of the measured quantity with respect to an input: 1 for level and density, ½ for flow through √ΔP.
β (beta ratio)
Ratio of orifice bore to pipe internal diameter, d/D.
Expansibility factor ε
Correction for gas compressibility across the orifice; exactly 1 for liquids, so its uncertainty is genuinely zero there.